338
6000 Series Oscilloscope User’s Guide
8
Reference
Waveform Data Header
A waveform may have more than one data set. Each waveform
data set will have a waveform data header. The waveform data
header consists of information about the waveform data set.
This header is stored immediately before the data set.
Waveform Data Header Size
A 32-bit integer that is the size of
the waveform data header.
Buffer Type
A 16-bit short that is the type of waveform data
stored in the file:
•
0 = Unknown data.
•
1 = Normal 32-bit float data.
•
2 = Maximum float data.
•
3 = Minimum float data.
•
4 = Not used in 6000 Series oscilloscopes.
•
5 = Not used in 6000 Series oscilloscopes.
•
6 = Digital unsigned 8-bit char data (for digital channels).
Bytes Per Point
A 16-bit short that is the number of bytes per
data point.
Buffer Size
A 32-bit integer that is the size of the buffer
required to hold the data points.
Example Program for Reading Binary Data
To find an example program for reading binary data, point your
web browser to
, select Library,
and click “Example Program for Reading Binary Data”.
Summary of Contents for InfiniiVision DSO6014L
Page 2: ...Agilent Technologies Agilent 6000 Series Oscilloscopes User s Guide...
Page 59: ...58 6000 Series Oscilloscope User s Guide 1 Getting Started...
Page 109: ...108 6000 Series Oscilloscope User s Guide 2 Front Panel Controls...
Page 125: ...124 6000 Series Oscilloscope User s Guide 3 Viewing and Measuring Digital Signals...
Page 251: ...250 6000 Series Oscilloscope User s Guide 5 Making Measurements...
Page 287: ...286 6000 Series Oscilloscope User s Guide 6 Displaying Data Interpreting Decoded LIN Data...
Page 321: ...320 6000 Series Oscilloscope User s Guide 7 Saving and Printing Data...
Page 370: ......