background image

+3$¶$3HUIRUPDQFH7HVW

HP 81130A/’31A Performance Test

12. Check the HP 81130A width at the following settings:

127(

Repeat the entire test for the second channel, if it is installed

Period

Width

Acceptable Range

TR Entry

5.882 

µ

s

5.882 

µ

s

µ

s

µ

s

0.9997 

µ

s   to  1.0003 

µ

s

4.9993 

µ

s   to  5.0007 

µ

s

2 - 7
2 - 8

Summary of Contents for HP 81130A

Page 1: ...3 3HUIRUPDQFH 7HVW...

Page 2: ...e label is shown on the display and which can vary according to the job that the button is doing 217 18286 38 6 6 This is an option shown on the dis play and is selected by use of the vernier keys It...

Page 3: ...If Channel 2 has been fitted to your instrument make an extra copy of the Test Results tables for entry of the results of tests on that channel In this case however it is not necessary to repeat the...

Page 4: ...0 bit vertical resolution Histogram Oscilloscope HP 54750A HP 54751A 20 GHz 15 bit vertical resolution Histogram Counter or HP 53132A 001 010 030 Frequency measurements 150 MHz High Stability Timebase...

Page 5: ...the 8710 1582 torque wrench to tighten and loosen SMA connectors This will ensure that the connectors are at the correct tightness and give the best signal transfer 50 Feedthrough Termination HP 1010...

Page 6: ...case The case must provide shielding and maintain grounding integrity 50 Ohm 0 1 10 W Feedthrough Termination The following parts are required 1 R1 53 6 1 10 W HP Part Number 0699 0146 2 R2 200 10 0 5...

Page 7: ...are fitted in the instrument When the HP 81130A is being tested therefore different situations can arise depending on whether you have a standard instrument or one that has had additional boards fitt...

Page 8: ...qty 2 of HP 81131A The LEVELS Screen in an HP 81130A with qty 2 of HP 81131A 7 0 1 3 77 51 9 6 02 75 HOD 2 V 2 3HULRG UHTXHQF 02 SV LGWK HDG 7UDL QV QV HOD SV LGWK HDG 7UDL QV QV HDG 3HU HDG 2 2 21 2...

Page 9: ...HELP 6 5 The HP 81130A display lists the instrument s products and serial number The display on your instrument should look similar to this The serial number given for the 5 0 applies to the Mainframe...

Page 10: ...rs an illustration of the settings is shown Set up the HP 81130A as follows 1 Select 02 75 217 18286 38 6 6 2 Select 287387 if second channel is installed 6HSDUDWH 2XW 3 Select MORE 21 screen and set...

Page 11: ...PDQFH 7HVW HP 81130A 31A Performance Test 127 Set ups are given in all the tests for 287387 and 287387 If you are testing a single channel instrument set up the 287387 screen with the settings given f...

Page 12: ...50 coaxial BNC SMA BNC Adaptor 127 The HP 53132A counter is used in frequency mode to meet the MIL CAL A uncertainty requirements for TAR Test Accuracy Ratio 4 1 3URFHGXUH Connect the HP 81130A to th...

Page 13: ...0A and set up as follows The MODE TRG Screen Setup 6 On the HP 81130A set up 287387 and 287387 pages as follows Configuring Output 1 287387 3 77 51 287387 02 75 38 6 6 RQWLQRXV 6WDUWHG DWHG 02 217 182...

Page 14: ...ments with 2 output channels it is necessary to a Configure ERWK channels b You can switch OFF the channels that are not being tested 7 Set the Counter to measure the frequency at the choosen input 1...

Page 15: ...Range TR Entry 2 500 ns 10 00 ns 50 00 ns 100 ns 500 ns 1 s 5 882 s 400 000MHz 100 MHz 20 MHz 10 MHz 2 MHz 1 MHz 170 0 kHz 399 9600 MHz to 400 0400 MHz 99 990 MHz to 100 010 MHz 19 9980 MHz to 20 002...

Page 16: ...01 200 ps TXLSPHQW 1HHGHG Digitizing Oscilloscope with Accessories Counter Cable 50 coaxial BNC SMA BNC Adaptor 3URFHGXUH 1 Connect HP 81130A to the Scope as shown Connecting HP 81130A to the Scope 2...

Page 17: ...87 pages as shown in the following illustrations Configuring Output Screen 1 Configuring Output Screen 2 3HU QV 21 287387 3 77 51 287387 02 75 HOD SV HDG 7UDL QV HDG 2IIVHW PSOLW P9 9 QV 02 LGWK QV 2...

Page 18: ...54121T Press AUTOSCALE Select the Display menu and set the Number of Averages to 32 Select the delta V menu and turn the voltage markers On Set the preset levels to 50 50 and press AUTO LEVEL SET Sel...

Page 19: ...idth Acceptable Range TR Entry 1 ns div 2 ns div 10 ns div 20 ns 100 ns 200 ns 200 ns 200 ns 1 s 1 s 1 250 ns 10 00 ns 50 00 ns 100 0 ns 500 0 ns 1 049875 ns to 1 450125 ns 9 799 ns to 10 201 ns 49 79...

Page 20: ...e Test 12 Check the HP 81130A width at the following settings 127 Repeat the entire test for the second channel if it is installed Period Width Acceptable Range TR Entry 5 882 s 5 882 s 1 s 5 s 0 9997...

Page 21: ...ter Cable 50 coaxial BNC SMA BNC Adaptor 3URFHGXUH Connect HP 81130A to the Scope as shown Range Variable Delay 0 ns to 3 00 s Fixed typical Delay of CLK REF IN to TRIGGER OUT 21ns CLK REF IN to OUTPU...

Page 22: ...81130A set up 287387 and 287387 pages as shown in the following illustrations Configuring Output Screen 1 SMA Cable 3 81 5 7 67 3 7 URQWHQG OUT 1 INPUT 1 2 3 4 TRIG 32 5 63 77 5 3 SMA SMA Adapter TRIG...

Page 23: ...t being tested If you then test the other channel c Switch ON the channel you are testing and switch OFF the other channel 3 Set the Digitizing Oscilloscope HP 54121T Press AUTOSCALE Set timebase to T...

Page 24: ...Check the HP 81130A delay at the following settings Oscilloscope Delay to min 16 0000 ns 127 With each new Oscilloscope Timebase setting set the HP 81130A Delay to 0ps press START ON EDGE POS1 and PR...

Page 25: ...the following setting 127 Subtract the fixed delay from the other readings 127 Repeat the entire test for the second channel if it is installed FUNCTION TI INPUT A INPUT B A B 50 50 3 81 5 7 67 TRIG...

Page 26: ...wing tests are required 1 Period Jitter 2 Width Jitter 3 Delay Jitter 7HVW 3HULRG LWWHU 7HVW 6SHFLILFDWLRQV RMS Jitter 0 001 15 ps TXLSPHQW 1HHGHG Digitizing Oscilloscope with Accessories Delay Line 2...

Page 27: ...sing the HP 54750A 54751A Equipment Set up using the HP 54121T SMA Cable 3 81 5 7 67 3 OUT 1 6 dB Attenuator SMA Cable TRIG INPUT 3 HOD LQH 32 5 63 77 5 3 OUTPUT SMA Cable 3 81 5 7 67 3 7 URQWHQG OUT...

Page 28: ...A set up 287387 and 287387 pages as shown in the following illustrations Configuring Output Screen 1 Configuring Output Screen 2 3HU 21 287387 3 77 51 287387 02 75 QV HOD SV LGWK HDG 7UDL QV QV HDG 2I...

Page 29: ...TOSCALE Select the Display menu and set the Number of Averages to 64 Select the Channel menu and set the Attenuation factor of channel 2 to 2 Set the VOLTS DIV of channel 2 to 10 mV div Set OFFSET to...

Page 30: ...g edge of the signal 7 Press MORE and HISTOGRAM Select the Window submenu and set Source is channel 2 Choose the Time Histogram Press WINDOW MARKER 1 and set it to 490 mV Press WINDOW MARKER 2 and set...

Page 31: ...A Performance Test 12 The RMS jitter for period of 20 ns is 15 2 ps Enter the result in the Test Report as TR entry 4 1 1 127 See the HP54750A User s Guide Service Guide to get the info needed to do t...

Page 32: ...scilloscope with Accessories Delay Line 22 ns Power Splitter 3URFHGXUH 1 Connect HP 81130A to the Scope as shown Equipment Set up for Jitter Test 2 Set up the HP 81130A as described in Initial Setup o...

Page 33: ...es as shown in the following illustrations Configuring Output Screen 1 Configuring Output Screen 2 3HU V 21 287387 3 77 51 287387 02 75 HOD SV LGWK HDG 7UDL QV QV HDG 2IIVHW PSOLW P9 9 QV 02 2 3HU V 2...

Page 34: ...TOSCALE Select the Display menu and set the Number of Averages to 128 Select the Channel menu and set the Attenuation factor of channel 2 to 2 Set the VOLTS DIV of channel 2 to 10 mV div Set OFFSET to...

Page 35: ...the first negative going edge of the signal Delay approx 80 ns 8 Press MORE and HISTOGRAM 9 Select the Window submenu and set Source is channel 2 Choose the Time Histogram Press WINDOW MARKER 1 and s...

Page 36: ...s TR entry 4 2 1 15 Set the HP 81130A for pulse width of 500ns 16 Repeat steps 7 to 13 127 TIME DIV 100ps div Approximate delay 530 ns 17 The RMS jitter for pulse width of 500 ns is 20 ps Enter the re...

Page 37: ...Oscilloscope with Accessories 3URFHGXUH 1 Connect HP 81130A to the Scope as shown Equipment Set up for Delay Jitter Test 2 For calculating the RMS jitter the rise time of the reference signal within a...

Page 38: ...n Initial Setup of the HP 81130A 4 On the HP 81130A press MORE and set up 287387 and 287387 pages as shown in the following illustrations Configuring Output Screen 1 Configuring Output Screen 2 3HU V...

Page 39: ...igitizing Oscilloscope HP 54121T Press AUTOSCALE Select the Display menu and set the Number of Averages to 64 Set the VOLTS DIV 10 mV div Set OFFSET to 500 mV Select the Timebase menu and set the TIME...

Page 40: ...follows 13 The RMS jitter for pulse width of 50 ns is 15 5 ps Enter the result in the Test Report as TR entry 4 3 1 14 Set HP 81130A for delay of 500 ns 15 Repeat steps 9 to 12 127 TIME DIV 100 ps di...

Page 41: ...V TXLSPHQW 1HHGHG 1 Digitizing Voltmeter DVM 2 50 Feedthrough Termination 0 1 10 W Adaptor 3 BNC to dual banana plug HP 1251 2277 4 SMA Cable SMA BNC Adaptor 3URFHGXUH Connect HP 81130A to the DVM as...

Page 42: ...l Setup of the HP 81130A 2 On the HP 81130A press MORE and set up 02 75 page as shown in the following illustration Configuring MODE TRIG Screen 3 81 5 7 67 3 90 50 OHM Feedthrough OUT 1 0 1 10W BNC D...

Page 43: ...er key OHIW highlight the of Segment and press ENTER Configuring Pattern Screen 4 On the HP 81130A set up 287387 and 287387 pages as shown in the following illustrations 6HJPHQW HQJKW RRSFQW 8SGDWH 28...

Page 44: ...t Screen 2 127 When you are testing instruments with 2 output channels it is necessary to a Configure ERWK channels UHT N 21 287387 3 77 51 287387 02 75 HOD LGWK 9 LJK RZ SV P9 9 02 2 HDG 7UDL QV HDG...

Page 45: ...high level at the following high level settings with the low level set to 0 0 V The low level may vary within 2 of amplitude 50 mV To check the low level change on the 3DWWHUQ page CH1 to Low Seg 7HVW...

Page 46: ...ess MORE and set up 287387 and 287387 pages as shown in the following illustrations Configuring Output Screen 1 6HJPHQW 8SGDWH 287387 3 77 51 287387 02 75 GGUHVV RZ HQJWK 02 LJK RWK DWD 6HJ LJK 6HJ RZ...

Page 47: ...st the other channel c Switch ON the channel you are testing and switch OFF the other channel 4 Check the HP 81130A low level at the following low level set tings with the high level set to 0 0 V UHT...

Page 48: ...0A 31A Performance Test The high level 0 0 V may vary 2 of amplitude 50 mV To check the low level change on the the 3DWWHUQ page CH1 to High Seg 127 Repeat the High and Low Level tests for the second...

Page 49: ...1130A MHz Pulse Generator Serial No ________________ Options ____________ Ambient temperature _______ C ____________ Relative humidity ________ ____________ ____________ Firmware Rev _____________ Lin...

Page 50: ...______________________________ 6 _________________________________________________ 7 _________________________________________________ 8 _________________________________________________ 9 ___________...

Page 51: ..._______________ ____________________________________________________ ____________________________________________________ UHTXHQF Counter Uncertainty factor BBBBBBBBBBBB BBBBBBBBBBBBBBBBBBBBBBBBBBBBBB...

Page 52: ...___ 1 0001 MHz __ __ 1 7 170kHz 169 983 kHz _______ 170 017 kHz __ __ 3HULRG LWWHU Scope Uncertainty factor ____________ BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB 75 QWU 7HVW LPLW FWXDO LP...

Page 53: ...99ns ________ 10 201 ns ____ ____ 2 3 50 0 ns 49 795 ns ________ 50 205 ns ____ ____ 2 4 100 ns 99 790 ns ________ 100 210 ns ____ ____ 2 5 500 ns 499 750 ns ________ 500 250 ns ____ ____ 2 6 1 s 0 99...

Page 54: ...s _________ 10 101ns ____ ____ 3 3 50 00 ns 49 895 n _________ 50 105 ns ____ ____ 3 4 100 0 ns 99 890 ns __________ 100 110 ns ____ ____ 3 5 500 0 ns 499 850 ns __________ 500 150 ns ____ ____ 3 6 3...

Page 55: ...93 V __________ 1 07 V ____ ____ 5 1 3 0 5 V 440 mV __________ 560 mV ____ ____ 5 1 4 0 1 V 48 mV __________ 152 mV ____ ____ RZ HYHO BBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBBB 75 QWU 7HVW...

Page 56: ...3 3HUIRUPDQFH 7HVW HP 81130A 31A Performance Test...

Reviews: