Vacuum Products Division
Manual 699910005, Hot Ion Source Replacement
Page 6
2
2
2
2
Filament Installation: Agilent Helium Leak
Filament Installation: Agilent Helium Leak
Filament Installation: Agilent Helium Leak
Filament Installation: Agilent Helium Leak
Detector Model G8610, G8611, or G8612
Detector Model G8610, G8611, or G8612
Detector Model G8610, G8611, or G8612
Detector Model G8610, G8611, or G8612
The same filament replacement kit is compatible with any model G8610, G8611, or G8612, as well
as legacy Agilent model VS series leak detectors.
2.1
Equipment Required for G8610, G8611, or G8612
The following tools are suggested for assembly.
Table 2-1. Equipment required for G8610, G8611, or G8612 models
Item
Item
Item
Item
Note
Note
Note
Note
Hot Ion Source Replacement Kit (VSFLDHIS)
--
Slotted screw driver
Use to remove leak detector covers
Metric Allen wrench set (Range 1.5 mm to
10 mm
--
Needle nose pliers or tweezer
--
M3 Phillips head screw driver
--
Torque wrench (must be adjustable to 40 in-
lbs (4.5 N-m) and 90 in-lbs (10.2 N-m))
--
Digital multimeter (Fluke 187 or equivalent)
For resistance and continuity measurements
NOTE
Inspect the new ion source kit for deformation, damage and filament
coating flaking prior to installation. Do not install if damaged. Touching
the filament under any condition will cause damage.
2.2
Installation Procedure for G8610, G8611, or G8612
For clarity, some items have been omitted from views.
Figure 2-1. Covers
Spectrometer
Header