164
5977 Series MSD Operation Manual
6
CI Maintenance
To Assemble the CI Ion Source
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Hex ball driver, 1.5 mm (8710-1570)
•
Hex ball driver, 2.0 mm (8710-1804)
•
Wrench, open-end, 10 mm (8710-2353)
Procedure
1
Assemble the ion focus lens, entrance lens, and lens insulator.
2
Slide the drawout plate and the drawout cylinder into the source body
(
3
Slide the assembled lens parts into the source body.
4
Install the setscrew that holds the lenses in place.
5
Install the repeller, repeller insulators, washer, repeller nut and source
heater block onto the source body.
6
Reinstall the filaments using the gold plated screws and spring washers.
C A U T I O N
Do not overtighten the repeller nut or the ceramic repeller insulators will break when
the source heats up. The nut should only be finger-tight.