General Maintenance
5
5977 Series MSD Operation Manual
129
4
If you are using an extractor EI ion source, disconnect the eight wires from
the ion source. Do not bend the wires any more than necessary (See
).
5
Trace the wires for the ion source heater and temperature sensor to the
feedthrough board. Disconnect them there.
6
Remove the thumbscrews that hold the ion source in place.
7
Pull the ion source out of the source radiator.
Table 23
Extractor EI ion source wires
Wire color
Connects to
Number of leads
Blue
Entrance lens
1
Orange
Ion focus
1
White
Filament 1 (top filament)
2
Red
Repeller
1
Black
Filament 2 (bottom filament)
2
Brown
Extractor lens
1
WA R N I N G
The analyzer operates at high temperatures. Do not touch any part until you are sure
it is cool.