5977 Series MSD Operation Manual
3
Operating in Electron Ionization (EI) Mode
To Perform High-Mass Testing (5977 Series MSDs)
Materials needed
•
PFHT calibration sample (5188-5357)
Procedure
1
Load tune file ATUNE.U then auto tune the MSD. See
2
Resolve the PFHT.M method under x\5977\PFHT.M where x is the
instrument number being used.
3
Update and save the method.
4
Load the PFHT calibration sample into a vial and place in position 2.
5
In
Instrument Control
view select
High Mass Check
from the
Checkout
menu.
6
Follow the instructions on screen.
7
The Run is completed and results are printed within 5 minutes. See