Installation Instructions
Agilent G3440-68011 Micro Electron Capture (µECD) 3rd Detector
Figure 8
MSD heated zone cable included in main communication grommet
9
Route the MSD heated zone cable behind the inlet fan and down the left
side of the GC. Include it in the grommet as shown in
Figure 10
on page 10.
10
Connect the third detector heated zone cable to the open end of the MSD
heated zone cable (on the side of the GC). See
Figure 10
on page 10.
11
Connect the
EPC5
female connector of the third detector communications
cable to the matching
EPC5
connector on the main communications harness.
The
EPC5
connector is typically located in EPC bay 5. See
Figure 9
on
page 9. You may need to lift any PCM or Aux EPC module to access
EPC5
on
the main harness.
12
Connect the
Aux Det1
female connector of the third detector
communications cable to the matching
Aux Det1
connector on the main
communications harness.
Aux Det1
is located in the back inlet EPC bay.
As an example, shown in
, is a TCD EPC communications cable
connection and routing.
MSD heated zone
cable
Grommet