Remove the detector cover by raising the cover vertically (until it is
perpendicular to the GC oven) and then firmly lifting up on the right side of
the cover to free it from the hinge pin. Slide the cover’s hinge pin out of the
hole on the inlet side and set the cover aside.
8
Remove the electronics cover. Loosen the screw on the left side of the
electronics cover. Open the cover 90 degrees and lift up on the right side.
Slide the cover off of the left side hinge.
9
Remove the GC right side cover. Loosen the single top screw, then slide the
cover towards the back of the GC and remove.
10
Remove the top back cover by loosening the four screws. Slide the top back
cover to the left until free, gently tilting out the lower end of the cover as
needed for removal.
11
Remove the GC left side cover by loosening the top screw, sliding the panel
towards the back, and lifting off the cover.
12
Put on an ESD wrist strap, and attach the ground to the GC sheet metal
frame for electrostatic protection.
Install the solenoid valve bracket and cables
1
Skip to
if the solenoid valve bracket is already installed on your GC.
2
Insert the four smaller valve plugs and two larger aux heater plugs into the
slots in the solenoid valve bracket. Match the labels on each plug to its
correct location as shown in
Figure 5
.
Figure 5
Connector installation for the solenoid valve bracket
3
Connect the aux zone/valve box cable connector into the analog and power
(A&P) board at P22. See
Figure 6
.
A1
V1 V2 V3 V4
A2