Installation and Operation Guide
3
Splitter Configurations
Disconnect tubing from the splitter
Loosen and remove the internal nut from the splitter fitting. Usually the
tubing and ferrule will fall out of the fitting.
Figure 15
Detector and column connections
To makeup
gas supply
Column in
Restrictor to
detector 1
Restrictor to
detector 2
Restrictor to
detector 3
C A U T I O N
Do not overtighten the fittings. One-eighth of a turn (about 15° clockwise from
finger-tight) is usually enough.