Splitter Configurations
3
Installation and Operation Guide
33
Maximum and minimum flows
The maximum suggested flow for MSDs depends on the vacuum pump used.
For diffusion pump and standard turbo systems, the flow should not exceed
2 mL/min. For performance turbo systems, the flow should not exceed
4 mL/min. These flow limits restrict the column flows and split ratios that can
be used with MSDs.
Try to have the flow through each restrictor tube at least equal to the
suggested minimum flow in
. Restrictors that have less flow will still
work, but peak broadening and/or tailing may result.
1
The makeup flow is listed in cell B 36 of the effluent splitter calculator. You
should have at least 0.5 mL/min for stable pressure regulation. Note that
this value will decrease as the oven temperature programs up.
2
Use the
Column Pressure/Flow Calculator
to determine the flow through each
restrictor at the maximum oven temperature of the method, add them and
subtract the calculated column flow at that temperature. This value should
be greater than 0.5 mL/min.
Table 4
Suggested minimum restrictor flows
Minimum carrier gas flow, mL/min
Restrictor internal
diameter, mm
Helium
Hydrogen
Nitrogen
Argon
0.10
0.400
0.500
0.125
0.110
0.18
0.720
0.900
0.225
0.198
0.20
0.800
1.000
0.250
0.220
0.25
1.000
1.250
0.313
0.275
0.32
1.280
1.600
0.400
0.352
0.45
1.800
2.250
0.563
0.495
0.53
2.120
2.650
0.663
0.583
Summary of Contents for G3183B
Page 1: ...Agilent Technologies Agilent G3183B Three Way Splitter Kit Installation and Operation Guide...
Page 4: ...4 Installation and Operation Guide...
Page 38: ...38 Installation and Operation Guide 3 Splitter Configurations...
Page 55: ......
Page 56: ...Agilent Technologies Agilent Technologies Inc Printed in USA April 2006 G3183 90120...