Agilent 1100 Series Fraction Collectors User’s Guide
75
Repairing the Fraction Collector
3
Simple Repairs
2 Remove the internal tray by pushing down
the plastic holder that holds it in position
underneath the metal latch (1) and sliding
the tray to the left at the same time (2).
3 Remove the corrugated waste tubing from
the front of the instrument (1) and slide the
internal tray’s waste tubing to the rear of the
outlet (2) the before sliding the tray out.
4 Install the new tray by sliding it into position
underneath the metal latch that holds it. The
waste tubing from the internal tray should be
guided as shown.
5 Make sure that the waste tubing is slid all the
way through the outlet (1). Its end should be
over the edge and below the level of the
laboratory desk that the system stands on to
avoid any back flow of solvent. Re-install the
corrugated waste tubing (2).
On completion of this procedure:
•
Re-install the tray(s) in the tray base.
•
Start the instrument.
•
Close the front cover.