Agilent 1100 Series Fraction Collectors User’s Guide
59
Repairing the Fraction Collector
3
Simple Repairs
Replacing the Inlet/Waste Tubings
Frequency
When contaminated, worn out or visibly damaged
Typically once every year
Tools required
None
Parts Required
Inlet / waste tubing assembly included in tubing kit
preparative scale 0.8 mm
ID,
PN
G1364-68711
or
Inlet / waste tubing assembly included in tubing kit
analytical scale 0.25 mm ID,
PN
G1364-68712
WA R N I N G
To avoid personal injury, keep fingers away from the needle area during fraction
collector operation.
WA R N I N G
Follow the described installation procedure exactly to maximize the lifetime of the
inlet / waste tubings and to avoid potential spills or fraction losses. Regularly
inspect the tubings and exchange them if they are worn out or show visible signs of
damage.