Agilent 1100 Series Fraction Collectors User’s Guide
15
Configuration and Operation of the Fraction Collector
1
Delay volumes and delay calibration
Delay volumes and delay calibration
Handling of Delay times and volumes
shows a schematic drawing of the fraction-collection part of the
Agilent 1100 series purification system with the two delay volumes V
D1
and
V
D2
. For peak-based fraction collection the system delay times t
D1
and t
D2
can
be calculated by dividing the delay volumes by the flow rate .
The delay volume V
D2
is a system constant and is 23 µl for the fraction
collector AS and 120 µl for the fraction collector PS. Delay volume V
D1
, which
is displayed in the
Fraction Collector Configuration
window, is determined
using the
Delay Volume Calibration
feature of the ChemStation software.
When a peak is detected during a purification run (
) the diverter valve
is triggered using the following delay time calculations:
Start of fraction collection: t = t
0
+ t
D1
End of fraction collection: t = t
E
+ t
D1
+ t
D2
Figure 2
Delay Volumes and delay times
ν
·