45
Instrument Status Indicator
The instrument status indicator indicates one of four possible instrument
conditions:
•
When the status indicator is
OFF
(and power switch light is on), the
instrument is in a
prerun
condition, and is ready to begin an analysis.
•
A
green
status indicator, indicates the instrument is performing an analysis
(
run
mode).
•
A
yellow
indicator indicates a
not-ready
condition. The instrument is in a
not-ready state when it is waiting for a specific condition to be reached or
completed (for example, front cover not installed), or while a self-test
procedure is running.
•
An
error
condition is indicated when the status indicator is
red
. An error
condition indicates the instrument has detected an internal problem which
affects correct operation of the instrument. Usually, an error condition
requires attention (for example, leak, defective internal components). An
error condition always interrupts the analysis.