Maintenance
7
Simple Repairs
Stator Face
When
Poor injection-volume reproducibility
Leaking injection valve
Tools required
•
1/4 inch wrench (supplied in accessory kit)
•
Hex key, 9/64 inch (supplied in accessory kit)
Parts required
#
Part number
Description
1
0100-1851
Stator face for G1329A
1
No stator face for G1329B (functionality part of stator head).
1
0101-1268
Stator face for G2260A
Preparations
•
Remove the front cover.
•
Remove the leak tubings (if necessary).
C A U T I O N
Removing the stator head
The stator face is held in place by the stator head. When you remove the stator head,
the stator face can fall out of the valve.
➔
Carefully handle the valve to prevent damage to the stator face
1
Remove all capillary fittings from the injection-valve ports
.
2
Loosen each fixing bolt two turns at a time. Remove the
bolts from the head.