31
Site Requirements and Specifications
2
Performance Specifications
Table 8
Performance Specifications Agilent 1260 Infinity Preparative Autosampler
(G2260A)
Type
Specification
Pressure
Operating range 0 – 40 MPa (0 – 400 bar, 0 – 5800psi)
GLP features
Early maintenance feedback (EMF), electronic records of
maintenance and errors
Communications
Controller-area network (CAN). GPIB (IEEE-448), RS232C,
APG-remote standard, optional four external contact closures and
BCD vial number output
Safety features
Leak detection and safe leak handling, low voltages in maintenance
areas, error detection and display
Injection range
0.1 – 900 µl in 0.1 µl increments (recommended 1 µl increments)
Up to 1800 µl with multiple draw (hardware modification required)
Up to 5000 µl with multiple draw (hardware modification required)
Replicate injections
1 – 99 from one vial
Precision
Typically < 0.5% RSD of peak areas from 5 – 2000 µl, Typically
< 1% RSD of peak areas from 2000 – 5000 µl, Typically < 3% RSD of
peak areas from 1 – 5 µl
Minimum sample volume
1 µl from 5 µl sample in 100 µl microvial, or 1 µl from 10 µl sample in
300 µl microvial
Sample viscosity range
0.2 – 50 cp
Sample capacity
100 × 2-ml vials in 1 tray
15 × 6-ml vials in ½ tray (Agilent vials only)
Injection cycle time
Typically 50 s, depending on draw speed and injection volume
Summary of Contents for G1329A
Page 1: ...Agilent 1260 Infinity Preparative Autosampler User Manual...
Page 6: ...6 Contents...
Page 22: ...22 1 Introduction to the Autosampler Agilent 1200 Infinity Interfaces...
Page 32: ...32 2 Site Requirements and Specifications Performance Specifications...
Page 52: ...52 3 Installing the Autosampler Transporting the Autosampler...
Page 74: ...74 6 Troubleshooting and Diagnostics Overview of the Sampler s Indicators and Test Functions...
Page 136: ...136 9 Identifying Cables RS 232 Cables...
Page 149: ...149 Index...