88
7
Maintenance
Simple Repairs
5
Reconnect the sample-loop fitting to the needle fitting.
6
Use
Needle Up
to lift the needle to a position
approximately 2mm above the seat.
7
Ensure the needle is aligned with the seat.
Next Steps:
8
On completion of this procedure: Install the front cover.
9
Select “
End
” in the maintenance function “
Change
Needle
”.