![Agilent Technologies G1329A User Manual Download Page 48](http://html1.mh-extra.com/html/agilent-technologies/g1329a/g1329a_user-manual_2867725048.webp)
48
3
Installing the Autosampler
Flow Connections
Flow Connections
1
Connect the pump outlet capillary to port 1 of the injection valve.
2
Connect column-compartment inlet capillary to port 6 of the injection
valve.
3
Connect the corrugated waste tube to the solvent waste from the leak plane.
4
Ensure that the waste tube is positioned inside the leak channel.
Parts required
#
Description
1
Parts from the Accessory kit
Preparations
•
Sampler is installed in the LC system
WA R N I N G
Toxic and hazardous solvents
The handling of solvents and reagents can hold health risks.
➔
When opening capillary or tube fittings solvents may leak out.
➔
Please observe appropriate safety procedures (for example, goggles, safety gloves
and protective clothing) as described in the material handling and safety data sheet
supplied by the solvent vendor, especially when toxic or hazardous solvents are
used.
N O T E
Do not extend the waste capillary of the autosampler. The siphoning effect might empty the
complete seat capillary introducing air into the system.
Summary of Contents for G1329A
Page 1: ...Agilent 1260 Infinity Preparative Autosampler User Manual...
Page 6: ...6 Contents...
Page 22: ...22 1 Introduction to the Autosampler Agilent 1200 Infinity Interfaces...
Page 32: ...32 2 Site Requirements and Specifications Performance Specifications...
Page 52: ...52 3 Installing the Autosampler Transporting the Autosampler...
Page 74: ...74 6 Troubleshooting and Diagnostics Overview of the Sampler s Indicators and Test Functions...
Page 136: ...136 9 Identifying Cables RS 232 Cables...
Page 149: ...149 Index...