18
Chapter 3
Making Measurements
Making Cable Fault Measurements
Figure 3-1
Calibrating for Cable Fault Location
* 50 ohm only. The 6 dB attenuators may be used to improve impedance matching.
** 75 ohm only.
NOTE
The distance-to-fault accessory kit, Option B7K, has all the 50 ohm accessories
required to make these measurements.
lists the cables and
accessories required for 50 ohm and 75 ohm measurements.
2. Configure the spectrum analyzer for the appropriate cable type.
a. Press the
Cable Type
menu key, and observe the Cable Type Selection
.
b. Page through available cable types using the
Tab
,
RPG
, or
Step
keys.
c. Press the
Select
menu key to select the appropriate cable type.
d. If you don’t see your particular cable type listed, you can modify one of the
“Custom” cables at the end of the cable type list. You will need to know the
Velocity Factor
and
Cable Loss per Metre
data for your
cable:
1. Select one of the unused Custom cables.
2. Press
Vel Factor
3. Key in the value
4. Press
Enter
5. Press
Cable Loss /m
6. Key in the value (dB).
7. Press
Enter
.
8. Press
Store Table
to save the changes.
Summary of Contents for ESA-E Series
Page 4: ...iv ...
Page 6: ...vi Contents ...
Page 7: ...1 1 Getting Started This chapter describes the process for installing the personality ...
Page 14: ...8 Chapter1 Getting Started Installing the Optional Measurement Personality ...
Page 16: ...10 Chapter2 Menu Maps and Key Descriptions Menus Menus Mode and Measure Menus ...
Page 21: ...15 3 Making Measurements ...
Page 44: ...38 Chapter4 If You Have a Problem Before You Call Agilent Technologies ...
Page 46: ...Index 2 Index ...