Chapter 3
25
Making Measurements
Making Return Loss Measurements
Figure 3-6
Return Loss Measurement Setup
* 50 ohm only. The 6 dB attenuators may be used to improve impedance matching.
** 75 ohm only.
NOTE
The coaxial load shown in
must be used if the DUT is a two-port
device.
3. Make the measurement.
a. Reconnect the device in place of the short/open.
b. Read the measurement and save it if required.
The return loss of the device is displayed on the screen. Use the marker to
evaluate the result and save the trace if required.
.