![Agilent Technologies ESA-E Series User Manual Download Page 34](http://html1.mh-extra.com/html/agilent-technologies/esa-e-series/esa-e-series_user-manual_2868110034.webp)
28
Chapter 3
Making Measurements
Making Return Loss Measurements
Where: RL is the measured return loss value.
VSWR is sometimes stated as a ratio. For example: 1.2:1 “one point two to one”
VSWR. The first number is the VSWR value taken from the table or calculated
using the formula. The second number is always 1.
VSWR
1
10
RL
–
20
----------
+
1
10
RL
–
20
----------
–
-----------------------
=