Chapter 3
19
Making Measurements
Making Cable Fault Measurements
Figure 3-2
Cable Type Selection Screen
3. Set up a maximum range value just greater than the length of the cable to be
tested:
a. Press the
Stop Distance
menu key.
b. Enter the appropriate value using the numeric key pad.
4. If required, set up a start distance offset to be used when displaying the trace
and calculating distances:
a. Press the
Start Distance
menu key.
b. Enter the appropriate value using the numeric key pad.
NOTE
The above steps may cause the trace-points parameter to be changed. To prevent
this from happening, change this parameter to manual before attempting steps 2
through 4. If you choose to override this parameter, the resultant trace may appear
discontinuous.
5. Calibrate the spectrum analyzer.
Calibration removes any errors introduced by the cabling and components of
the test setup before making the measurement.
NOTE
Press the
Esc
front panel key to cancel this procedure at any stage.
a. Disconnect the cable to be tested.
b. Press the
Meas Setup
front-panel key.
Summary of Contents for ESA-E Series
Page 4: ...iv ...
Page 6: ...vi Contents ...
Page 7: ...1 1 Getting Started This chapter describes the process for installing the personality ...
Page 14: ...8 Chapter1 Getting Started Installing the Optional Measurement Personality ...
Page 16: ...10 Chapter2 Menu Maps and Key Descriptions Menus Menus Mode and Measure Menus ...
Page 21: ...15 3 Making Measurements ...
Page 44: ...38 Chapter4 If You Have a Problem Before You Call Agilent Technologies ...
Page 46: ...Index 2 Index ...