Installation Note E8364-90019
Analyzers WITH Option UNL but WITHOUT Option 016
Refer to
for the following procedure. The new parts referenced in this procedure are
listed in
Install these cables
in the order listed
. Use a 5/16-inch torque wrench set to 10 in-lbs.
• W71 E8364-20177 REFERENCE 2 RCVR R2 IN to A29 channel R2 mixer
• W70 E8364-20176 REFERENCE 1 RCVR R1 IN to A28 channel R1 mixer
• W68 E8364-20044 A26 test port 1 coupler to PORT 2 CPLR ARM
• W72 E8364-20046 PORT 2 RCVR B IN to A30 channel B mixer
• W82 E8364-20054 A37 step attenuator to PORT 2 SOURCE OUT
• W84 E8364-20040 PORT 2 CPLR THRU to A39 bias tee
• W66 E8364-20048 Channel R2 detector to REFERENCE 2 SOURCE OUT
• W67 E8364-20043 A25 test port 1 coupler to PORT 1 CPLR ARM
• W69 E8364-20045 PORT 1 RCVR A IN to A27 channel A mixer
• W81 E8364-20053 A36 step attenuator to PORT 1 SOURCE OUT
• W83 E8364-20039 PORT 1 CPLR THRU to A38 bias tee
• W65 E8364-20047 Channel R1 detector to REFERENCE 1 SOURCE OUT
Figure 9
Cable Installation, Analyzers with Option UNL but without Option 016