22
Installation Note E8364-90019
Step 10. Enable Option 014
Procedure Requirements
• The analyzer must be powered up and operating to perform this procedure.
• The Network Analyzer program must be running.
• A mouse must be connected to the analyzer for this procedure.
Option Enable Procedure
1. On the analyzer’s
System
menu, point to
Service
, and then click
Option Enable
.
2. In the
Select Desired Option
list, click
014 - Configurable Test Set
.
3. Click
Enable
.
4. Click
Yes
in answer to the displayed question in the
Restart Analyzer?
box.
5. When the installation is complete, click
Exit
.
Option Verification Procedure
Once the analyzer has restarted and the Network Analyzer program is again running:
1. On the analyzer’s
Help
menu, click
About Network Analyzer
.
2. Verify that “014” is listed after “Options:” in the display.
3. Click
OK
to close this screen.
NOTE
If Options 014 has not been enabled, perform the
again. If the option is still not enabled, contact Agilent
Technologies. Refer to