User’s and Service Guide
19
E8257D Option HAR
Appendix
Appendix
Modifying Synthesizer Self Test Limits
This procedure allows you to modify the synthesizer A6 Frac N self test limits for test 704,
after replacing the A6 Frac N assembly. (Firmware versions C.04.70 and later).
Open a browser window and type the IP address of the instrument, followed by ‘/update’ as
shown: http://xxx.xxx.xxx.xxx/update.
1. Enter the following text in the “Manual Update” form as shown below.
ll "SYNTH", 14, 0.5
ul "SYNTH", 14, 13.5
ll "SYNTH", 15, 0.5
ul "SYNTH", 15, 13.5
ll "SYNTH", 16, 1.5
ul "SYNTH", 16, 13.5
ky 53004
2. Press
Execute
(next to the manual update form). When the update is complete
“Operation Completed” will be displayed.
3. Cycle the power on the instrument.
If you need to revert to the standard self test limits, use the following sequence:
ll "SYNTH", 14, 1.5
ul "SYNTH", 14, 13.5
ll "SYNTH", 15, 3.0
ul "SYNTH", 15, 13.5
ll "SYNTH", 16, 2.0
ul "SYNTH", 16, 13.5
ky 51873