User’s and Service Guide
3
E8257D Option HAR
Specifications
Specifications
The E8257D Option HAR provides improved absolute phase noise in the 0.98 to 500 MHz
frequency range. To achieve these improvements the Divider Mode must be turned On.
See
The E8257D Option HAR impacts standard instrument performance for frequencies
≤
3.2 GHz. Standard operation for frequencies above 3.2 GHz has not been affected. If
Divider Mode is left on for frequencies outside of 0.98 to 500 MHz the “ALC Bandwidth”
will be limited to 100 Hz.
Phase Noise
Divider Mode On
Divider Mode Off
Standard performance per the E8257D data sheet.
Maximum Level Output Power
Divider Mode On
Typical performance see
Divider Mode Off
Table 1
E8257D Option HAR Specifications
1
(25 °C)
1. +16 dBm
Frequency
1 Hz
10 Hz
100 Hz
1 kHz
10 kHz
100 kHz
10 MHz
−
93
−
120
−
142
−
152
−
158
−
158
100 MHz
−
73
−
100
−
122
−
140
−
145
−
145
Table 2
Maximum Power Specification Impacts
Frequency
Delta
250 kHz to 3.2 GHz
−
3 dB from Standard specification
>
3.2 GHz
no change from standard specifications