16
User’s and Service Guide
E8257D Option HAR
Service
100 MHz Phase Noise Test Limits
100 MHz Test Procedure
Using the E5504B Phase Noise Measurement System.
Step 1.
Select
Define
>
Limit Lines
.
Step 2.
Enter the Noise Spec data (
) under the
Upper Noise Limit
tab.
Step 3.
Select
Define
>
Measurement
.
Table 14
Phase Noise Test Limits
100 MHz
1 Hz
10 Hz
100 Hz
1 kHz
10 kHz
100 kHz
Noise
Spec
−
73
−
100
−
122
−
140
−
145
−
145