Performance Tests
Instrument Accuracy Test
2-10
Agilent E4418B/E4419B Service Guide
Table 2-3: Instrument Accuracy Results (For Indication Only)
Note
The nominal outputs for the 3
µ
W, 30
µ
W, 300
µ
W, 3 mW, 10 mW,
30 mW and, 100 mW settings are 3.16
µ
W, 31.6
µ
W, 316
µ
W,
3.156 mW, 10.03 mW, 31.78 mW, and 101.3 mW respectively.
Note
It is not necessary to check instrument accuracy in dBm. The
power meter uses the same internal circuitry to measure power
and mathematically converts watts to dBm.
Range
Calibrator
Setting
Min
1
1. These performance limits are determined by the zero set specification of the power sensor
used in the measurement plus the measurement noise.
Channel A
Actual Results
Channel B
Actual Results
(Agilent E4419
B only)
Max
1
3
µ
W
3.100
µ
W
_______________
_______________
3.230
µ
W
10 µ
W
9.900
µ
W
_______________
_______________
10.10
µ
W
30 µ
W
31.40
µ
W
_______________
_______________
31.80
µ
W
100 µ
W
99.50
µ
W
_______________
_______________
100.5
µ
W
300 µ
W
314.00
µ
W
_______________
_______________
318.00
µ
W
1 mW
0.995 mW
_______________
_______________
1.005 mW
3 mW
3.141 mW
_______________
_______________
3.171 mW
10 mW
9.984 mW
_______________
_______________
10.08 mW
30 mW
31.63 mW
_______________
_______________
31.94 mW
100 mW
100.9 mW
_______________
_______________
101.8 mW
4402serv.book Page 10 Tuesday, October 14, 2003 3:18 PM
Summary of Contents for E4418B
Page 19: ...1 Specifications 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 37: ...2 Performance Tests 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 69: ...3 Adjustments 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 81: ...4 Theory of Operation 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 95: ...5 Replaceable Parts 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 119: ...6 Troubleshooting 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...