Measurement Uncertainty Analysis – Instrument Accuracy Test
Measurement Introduction
Agilent E4416A/E4417A Service Guide
A-7
Worst case error on 15.8V maximum power setting
The relationship between the applied voltage from the Universal Source to the
corresponding power is given from:
1mW = 0.14493mV
This is a linear relationship, for example, 0.1mW is equal to 0.014493volts. This
relationship is valid between 0dBm and –30dBm.
However, above 0dBm a look up table is used for voltage to mW conversion. This is
due to the sensors non linear behavior at higher power levels. The voltage correction
at the appropriate points is shown in Table A-2.
Therefore to attain a power of 100mW as shown in the Table A-2, instead of applying
14.4493 Volts, the actual amount of voltage required from the Universal Source
would be:
14.4493 x 1.075533 = 15.6Volts
Voltage Error
Distribution
Standard Uncertainty
257.75uV
Rectangular
148.8uV
Table A-2: Voltage Correction Multiplier
Power (mW)
Voltage Correction Multiplier
1
1
10
1.006577
100
1.075533
Error = ((10ppm x 15.8V) + (0.3ppm x 100.0V)) + (0.5ppm x 15.8V) + 5((0.15ppm x 15.8V) + (0.1ppm x 100V))
= 257.75uV
Percentage Error = 0.00163%
4402serv.book Page 7 Tuesday, October 14, 2003 3:18 PM
Summary of Contents for E4418B
Page 19: ...1 Specifications 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 37: ...2 Performance Tests 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 69: ...3 Adjustments 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 81: ...4 Theory of Operation 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 95: ...5 Replaceable Parts 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 119: ...6 Troubleshooting 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...