Performance Tests
Instrument Accuracy Test
2-14
Agilent E4418B/E4419B Service Guide
14. Set the DC calibrator output to the values shown in Table 2-4. For
each setting, set the filter length of the power meter to the
corresponding value, and verify that the power meter reading is
within the limits shown. Refer to step 10 for details on how to
setup the filter length
.
Note
The test limits shown in this table do not take into account the
corresponding measurement uncertainty values.
Table 2-4: Instrument Accuracy Results
DC
Calibrator
Setting
Effective
Power
Applied
Power
Meter
Filter
Setting
Meas
Unc
Lower
Limit
Ch A
Result
Ch B
Result
Upper
Limit
458.31uV
3.1623 uW
256
0.028 uW
3.0977uW
3.2269uW
1.4493mV
10uW
256
0.027 uW
9.904 uW
10.096uW
4.5831mV
31.6228uW
256
0.027 uW
31.428uW
31.819uW
14.493mV
100uW
64
0.073 uW
99.49uW
100.51uW
45.831mV
316.2278uW
64
0.104 uW
314.72uW
317.74uW
144.93mV
1mW
16
0.2 uW
0.9954mW
1.0047mW
459.12mV
3.1623mW
16
0.9 uW
3.1477mW
3.1769mW
1.4588V
10mW
16
1.6 uW
9.954mW
10.046mW
4.6872V
31.6228mW
16
4.8 uW
31.477mW
31.769mW
15.588V
100mW
16
18.5 uW
99.54mW
100.46mW
4402serv.book Page 14 Tuesday, October 14, 2003 3:18 PM
Summary of Contents for E4418B
Page 19: ...1 Specifications 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 37: ...2 Performance Tests 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 69: ...3 Adjustments 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 81: ...4 Theory of Operation 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 95: ...5 Replaceable Parts 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 119: ...6 Troubleshooting 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...