5GVVKPI1HHUGVU
)GPGTCN2QYGT/GVGT(WPEVKQPU
'2/27UGT U)WKFG
7
Enter the offset using the
,
,
and
keys. Complete the entry
by pressing the
key.
Continue adding/editing values until you have entered all the data you
require.
8
When you have finished editing the table press
to save the table.
N O T E
If you measure a signal with a frequency outside the frequency range
defined in the frequency dependent offset table, the power meter uses the
highest or lowest frequency point in the frequency dependent offset table
to calculate the offset.
&QPG
E4416-90023.book Page 63 Thursday, August 9, 2001 9:17 AM