-113
Undefined header
A command was received that is not valid for this power meter. You
may have misspelled the command, it may not be a valid command or
you may have the wrong interface selected. If you are using the short
form of the command, remember that it may contain up to four letters.
For example,
TRIG:SOUR IMM
.
-121
Invalid character in number
An invalid character was found in the number specified for a
parameter value.
For example,
SENS:AVER:COUN 128#H
.
-123
Exponent too large
A numeric parameter was found whose exponent was larger than
32,000.
For example,
SENS:COUN 1E34000
.
-124
Too many digits
A numeric parameter was found whose mantissa contained more than
255 digits, excluding leading zeros.
-128
Numeric data not allowed
A numeric value was received within a command which does not
accept a numeric value.
For example,
MEM:CLE 24
.
-131
Invalid suffix
A suffix was incorrectly specified for a numeric parameter. You may
have misspelled the suffix.
For example,
SENS:FREQ 200KZ
.
-134
Suffix too long
A suffix used contained more than 12 characters.
For example,
SENS:FREQ 2MHZZZZZZZZZZZ
.