Press
and the channel
softkey to display the
(TGSWGPE[
pop up window.
(KIWTG (TGSWGPE[2QR7R9KPFQY
6
Use the
,
,
and
keys to select and change the digits to
match the frequency of the signal you want to measure.
7
Press
or
as required to complete the entry.
Now make the measurement.
8
Connect the power sensor to the signal to be measured.
9
The corrected measurement result is displayed.
N O T E
If the measurement frequency does not correspond directly to a
frequency in the sensor calibration table, the power meter calculates the
calibration factor using linear interpolation.
If you enter a frequency outside the frequency range defined in the sensor
calibration table, the power meter uses the highest or lowest frequency
point in the sensor calibration table to set the calibration factor.
N O T E
When
display mode is chosen, the frequency you entered
and the sensor table identifier is displayed in the upper window. Also,
pressing
displays the frequency you entered and calibration
factor for each channel derived from the selected sensor tables.
Frequency
Cal Fac
(TGS
)*\
/*\
5KPING0WOGTKE
Frequency
Cal Fac