Lists not same length
This occurs when
SENSe:CORRection:CSET[1]|CSET2:STATe
is
set to
ON
and the frequency and calibration/offset lists do not
correspond in length.
-230
Data corrupt or stale
This occurs when a
FETC?
is attempted and either a reset has been
received or the power meter state has changed such that the current
measurement is invalidated (for example, a change of frequency
setting or triggering conditions).
-230
Data corrupt or stale;Please zero and calibrate Channel A
When
CAL[1|2]:RCAL
is set to
ON
and the sensor currently
connected to channel A has not been zeroed and calibrated, then any
command which would normally return a measurement result (for
example
FETC?
,
READ?
, or
MEAS?
) will generate this error message.
-230
Data corrupt or stale;Please zero and calibrate Channel B
When
CAL[1|2]:RCAL
is set to
ON
and the sensor currently
connected to channel B has not been zeroed and calibrated, then any
command which would normally return a measurement result (for
example
FETC?
,
READ?
, or
MEAS?
) will generate this error message.
-230
Data corrupt or stale;Please zero Channel A
When
CAL[1|2]:RCAL
is set to
ON
and the sensor currently
connected to channel A has not been zeroed, then any command which
would normally return a measurement result (for example
FETC?
,
READ?
, or
MEAS?
) will generate this error message.
-230
Data corrupt or stale;Please zero Channel B
When
CAL[1|2]:RCAL
is set to
ON
and the sensor currently
connected to channel B has not been zeroed, then any command which
would normally return a measurement result (for example
FETC?
,
READ?
, or
MEAS?
) will generate this error message