RF Switch Error Messages 53
Appendix C
Appendix C
RF Switch Error Messages
The following error messages are unique to the E1470A. See the
appropriate command module or VXI Controller module manual for a more
complete list of possible error messages.
Error
Number
Message Generated
Description
Commands that
may cause error
-108
“Parameter not allowed”
A parameter was specified that is not valid for the
command. More than 80 channels specified in
command.
DIAG:CLOS; DIAG:CLOS?;
DIAG:OPEN; DIAG:OPEN?
-222
“Data out of Range”
Invalid numerical state parameter.
*RCL
*SAV
-240
“Hardware Error”
More than 60 msec was required for the relays to
settle.
*OPC
*OPC?
*WAI
2001
“Invalid channel number” Invalid channel number in <source> parameter
[ROUT:]PATH[:COMM]
[ROUT:]PATH[:COMM]?
2022
“Invalid relay number”
Invalid relay number in command
DIAG:CLOS; DIAG:CLOS?;
DIAG:OPEN; DIAG:OPEN?
2023
“Invalid common bank
number”
Invalid <comm> parameter in command.
[ROUT:]PATH[:COMM]
[ROUT:]PATH[:COMM]?
2024
“Invalid source bank
number”
Invalid source number in <source> parameter
[ROUT:]PATH[:COMM]
[ROUT:]PATH[:COMM]?
2025
“Invalid Common-Source
Combination”
Even though the <comm> and <source>
parameters are valid, the combination is not valid.
The specified source cannot connect to the
specified COM terminal.
[ROUT:]PATH[:COMM]
[ROUT:]PATH[:COMM]?
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com