Error Messages
143
Appendix C
Appendix C
Error Messages
Code
Message
Cause
-101
Invalid character
Unrecognized character in specified parameter.
-102
Syntax Error
Command is missing a space or comma between parameters.
-103
Invalid separator
Command parameters are not separated by a comma.
-104
Data Type Error
The wrong data type (i.e. number, character, string, expression) was used
when specifying a parameter.
-108
Parameter not allowed
Parameter specified in a command where none is allowed.
-109
Missing parameter
No parameter specified when required.
-113
Undefined header
Command header was incorrectly specified.
-124
Too many digits
>257 digits were specified for a parameter.
-128
Numeric data not allowed
A number was specified for a parameter when a letter is required.
-131
Invalid suffix
Parameter suffix incorrectly specified.
-138
Suffix not allowed
Parameter suffix is specified when one is not allowed.
-141
Invalid character data
The parameter type specified is not allowed.
-161
Invalid block data
Mismatch between character count in header and actual number of
characters.
-178
Expression data not allowed
A parameter is enclosed in parenthesis.
-221
Settings conflict
Digital I/O command settings are in conflict (e.g., control asserted when in
a handshake mode other than NONE).
-222
Data out of range
Value specified is out of the legal range for parameter.
-224
Illegal Parameter Value
Inconsistent parameter value or block not found.
-240
Hardware Error
Hardware error detected during power-on cycle.
Return Digital I/O Module to Agilent Technologies for repair.
-410
Query Interrupted
Data is not read from the output buffer before another command is issued.
-420
Query unterminated
Command which generates data not able to finish executing due to a
Digital I/O Module configuration error.
-430
Query deadlocked
Command execution cannot continue since the mainframe's command
input and data buffers are full. Clearing the instrument restores control.
+1000
Out of memory
No memory available.
+2006
Undefined command
Command not recognized by this instrument.
+2025
Invalid port number for access TYPE
The port number specified is not valid for the [:type] set.
+2026
Port number out of range
The port number specified is out of the legal range of port numbers.
+2027
Invalid bit number for access TYPE
The bit number specified is not valid for the port [:type] set.
+2028
LW64 & LW96 not supported by this card Attempt to replace [:type] with either LW64 or LW96 keywords.
+2029
Duplicate memory block name
The memory block name specified already exists.
+2030
Invalid number of bytes for TRACE
access TYPE
The number of bytes specified in a trace does not match the [:type] set.
Summary of Contents for E1330B
Page 2: ......
Page 10: ...8 Notes ...
Page 11: ...9 Notes ...
Page 12: ...10 Notes ...
Page 32: ...30 Configuring the Agilent E1330B Digital I O Chapter 2 Notes ...
Page 42: ...40 Using the Agilent E1330B Digital I O Module Chapter 3 Notes ...
Page 58: ...56 Understanding the Agilent E1330B Digital I O Module Chapter 4 Notes ...
Page 104: ...102 Agilent E1330B Digital I O Module Command Reference Chapter 5 Notes ...