Chapter 5: Troubleshooting
Display Trouble Isolation
83
Display Trouble Isolation
Display Trouble Isolation Flowchart
No Display Trouble
Isolation
Remove cabinet and
install fan guard
Check fan
connections and
power-on
Fans running?
Power LED lit?
Connect external
monitor, cycle power,
and check power up
sequence
Power-up
displayed on ext.
monitor
Replace Display
Board
Go to Primary
Trouble Isolation
Go to 'Power Supply
Trouble Isolation.'
Go to 'To check the
backlight Inverter
voltages.'
Display on
oscilloscope?
Yes
No
Yes
No
Replace
Motherboard
Assembly
No
No
A
B
C
D
E
54830F02a
Display on
oscilloscope?
No
Done.
Yes
Display on
oscilloscope?
Done.
Yes
No
Check the display
board video signals
Done.
Yes
No
Summary of Contents for DSO80204B
Page 7: ...6...
Page 17: ...Chapter 1 General Information Oscilloscope Dimensions 16...
Page 34: ...3 To run the self calibration 35 Calibration...
Page 35: ...34 Calibration This chapter provides self calibration procedures for the oscilloscope...
Page 92: ...Chapter 5 Troubleshooting Motherboard Verification 91 Figure 5 14 Sense cable...
Page 140: ...Chapter 7 Replaceable Parts Exploded Views 139 Exploded Views Front Frame and Front Panel...
Page 141: ...Chapter 7 Replaceable Parts Exploded Views 140 Fan and Acquisition Assembly...
Page 142: ...Chapter 7 Replaceable Parts Exploded Views 141 Power Supply and PC Motherboard...
Page 143: ...Chapter 7 Replaceable Parts Exploded Views 142 Sleeve and Accessory Pouch...
Page 149: ...Chapter 8 Theory of Operation 148 Figure 8 1 Oscilloscope Block Diagram...
Page 157: ...156...
Page 159: ...158 Index...