Chapter 6: Replacing Assemblies
To remove and replace the covers
105
10
Carefully slide the bottom cover off the frame while spreading the top sleeve open as
shown.
11
Turn the instrument over and carefully slide the top sleeve off of the frame.
12
To replace the covers, reverse the above procedure.
Be sure to keep ribbon cables out of the way when replacing the covers, particularly the flex
cable and connector for the AutoProbe assembly at the bottom front of the oscilloscope.
C A U T I O N
PROPERLY TIGHTEN HANDLE AND SCREWS!
Tighten the side handle screws to 2.4 Nm (21 in-lbs) and rear feet screws to 2 Nm (18 in-lbs).
Figure 6-2
Remove bottom cover
Summary of Contents for DSO80204B
Page 7: ...6...
Page 17: ...Chapter 1 General Information Oscilloscope Dimensions 16...
Page 34: ...3 To run the self calibration 35 Calibration...
Page 35: ...34 Calibration This chapter provides self calibration procedures for the oscilloscope...
Page 92: ...Chapter 5 Troubleshooting Motherboard Verification 91 Figure 5 14 Sense cable...
Page 140: ...Chapter 7 Replaceable Parts Exploded Views 139 Exploded Views Front Frame and Front Panel...
Page 141: ...Chapter 7 Replaceable Parts Exploded Views 140 Fan and Acquisition Assembly...
Page 142: ...Chapter 7 Replaceable Parts Exploded Views 141 Power Supply and PC Motherboard...
Page 143: ...Chapter 7 Replaceable Parts Exploded Views 142 Sleeve and Accessory Pouch...
Page 149: ...Chapter 8 Theory of Operation 148 Figure 8 1 Oscilloscope Block Diagram...
Page 157: ...156...
Page 159: ...158 Index...