5
Agilent DNA 7500 and DNA 12000 Kit
1
Table 1
Physical Specifications
Type
Specification
Analysis time
30 minutes
Samples per chip
12
Sample volume
1 µL
Kit stability
4 months (see box for storage temperatures)
Kit size
12 samples/chip = 300 samples/kit
Table 2
Analytical Specifications
Type
Agilent DNA 7500
Agilent DNA 12000
Sizing range
100 – 7500 bp
100 – 12000 bp
Sizing resolution
100 – 1000 bp: 5 %
1000 – 7500 bp: 15 %
100 – 1000 bp: 5 %
100 – 12000 bp: 15 %
Sizing accuracy
± 10 %
± 15 %
Sizing reproducability
5 % CV
5 % CV
Quantitation
accuracy
1
1
Determined using the respective DNA ladder as sample
20 %
25 %
Quantitation
reproducibility
1
100 – 1000 bp: 10 % CV
1000 – 7500 bp: 5 % CV
100 – 1000 bp: 15 % CV
1000 – 12000 bp: 10 % CV
Quantitative range
1
0.5 – 50 ng/µl
0.5 – 50 ng/µl
Maximum salt
250 mM for KCl
15 mM for MgCl
2
250 mM NaCl
250 mM for KCl
15 mM for MgCl
2
250 mM NaCl