35
2 Laboratory setup requirements
Electrical requirements
BenchBot Robot Safety and Installation Guide
Electrical requirements
Be sure to connect the BenchBot Robot to an AC circuit that is properly
grounded.
Related information
Requirement
Value
Voltage
100–240 VAC
Frequency
50/60 Hz
Current
4 A
Power consumption
200 W typical
Fuses
6.3 A, 250 V, time delay, 5 mm x 20 mm
Chassis plug
IEC 60320 C14
For information about...
See...
General placement considerations
“General placement considerations” on
page 12
Environmental requirements
“Environmental requirements” on
page 36
Labware and performance
specifications
“Labware and performance
specifications” on page 37
Computer requirements
Summary of Contents for BenchBot