Recommended starting gradients
Stationary phase
Usage notes
Most reversed-phase
columns
5% methanol or acetonitrile initially and 100%
methanol or acetonitrile as the final solvent.
ZORBAX Eclipse PAH*
30 or 40% acetonitrile initially, to 100%
acetonitrile as final solvent. It may be necessary
to cool column to 15 to 20 °C for improved
resolution.
ZORBAX Bonus-RP
Lower concentration of organic mobile phase
modifier is required for compound elution,
compared to traditional long-chain alkyl
stationary phases.
Large molecule
columns (for example,
Poroshell 300 or
ZORBAX 300SB
columns)
Separations of polypeptides, proteins, DNA
fragments, etc. often use trifluoroacetic or
formic acid modifier for pH control and/or as
an ion-pairing additive for desired retention
and selectivity. Organic buffers such as TRIS
are especially useful for intermediate pH
applications.
* Best column lifetime is achieved at <40 °C.
Summary of Contents for and C8
Page 12: ...ZORBAX Poroshell Pursuit Polaris agilent com chem columnchoices Agilent QC QC LC QC...
Page 14: ...10 20 C8 CN...
Page 15: ...QC Agilent pH pH pH Poroshell 120 ZORBAX RRHT RRHD Fast Guard agilent com chem fastguards...
Page 60: ...ZORBAX Poroshell Pursuit Polaris agilent com chem columnchoices Agilent QC QC LC QC LC...
Page 62: ...10 20 C8 CN QC...
Page 63: ...Agilent pH pH pH Poroshell 120 ZORBAX RRHT 1 8 um RRHD Fast Guard agilent com chem fastguards...
Page 67: ...agilent com chem fastguards 1 2 10 20 3 100 4 5 4 1 8 m...
Page 68: ...20 30 50 50 20 30 60 40 ACN 0 02 M 60 40 ACN H20...
Page 70: ...ZORBAX Pursuit Polaris agilent com chem columnchoices Agilent...
Page 72: ...10 20 C8 CN...
Page 78: ...6 100 100 100 1 8 50 50 20 30 20 30 H2O 60 40 60 40 0 02...