Parts Replacement
Remove the Connector Board
Released: MAY 2000
Agilent 35900E Service Manual
16
Remove the Connector Board
1.
Remove the top cover. (See page 13.)
2.
Remove the MIO bracket. (See page 13.)
3.
Remove the two screws attaching the connector board to the plate. The
board will separate from the plate. (See part "A" in Figure 5.)
Remove/Replace the Mother Board
1.
Remove the top cover. (See page 13.)
2.
Remove the front panel. (See page 14.)
3.
Remove the MIO bracket and connector board. (See page 15 and 16.)
4.
Disconnect the power supply cable. (See page 15.)
5.
Remove the two screws attaching the mother board to the back of the unit.
(See Figure 5.)
6.
Slide the mother board out the front of the unit. (See part "4" in Figure 5)
7.
When you are ready to replace the mother board, slide the new mother
board into the guides from the front of the unit so that the cable
connections fit through the back of the unit.
8.
Replace the two screws on the back of the unit and all other removed
parts.