Parts Replacement
Remove/Replace the Power Supply
Released: MAY 2000
Agilent 35900E Service Manual
15
Remove/Replace the Power Supply
1.
Remove the top cover. (See page 13.)
2.
Remove the screw attaching the power supply to the right side of the
frame. (See Figure 4.)
3.
Disconnect the power supply cable. (See Figure 5.)
a.
Slide a flat-blade screwdriver (or your thumb nail) between the back
of the connector and the white locking ridge.
b.
Grasp the ends of the connector between your thumb and forefinger
and pull firmly upwards.
4.
Pull the power supply unit forward to slip it off of the tabs, then pull
upwards to remove it from the frame.
5.
When you are ready to replace the power supply, fit the new power supply
over the hooked tabs and press it against the frame.
6.
Reattach the power supply to the frame by replacing the holding screw on
the right side of the unit.
7.
Reconnect the power supply cable to the white connector. Push the
connector down firmly, until the locking ridge fits over the back of the
connector.
Remove the MIO Bracket
1.
Remove the top cover. (See page 13.)
2.
Remove the MIO card by first removing the two thumb screws holding the
card at the back of the unit. The card can then be pulled out from the rear
plate.
3.
Remove the three screws attaching the MIO bracket to the frame. (See
part "A" in Figure 5.)
4.
Firmly pull up on both the plate and the connector board, holding them
together at a right angle.
Summary of Contents for Agilent 35900E
Page 1: ...Agilent 35900E Dual Channel Interface Service Manual ...
Page 2: ...35900E Dual Channel Interface Agilent 35900E Service Manual ...
Page 5: ...Released MAY 2000 Agilent 35900E Service Manual 5 1 Introduction ...
Page 7: ...Released MAY 2000 Agilent 35900E Service Manual 7 2 Troubleshooting the 35900E ...
Page 10: ...Released MAY 2000 Agilent 35900E Service Manual 10 3 Parts Replacement ...
Page 18: ...Released MAY 2000 Agilent 35900E Service Manual 18 4 Instrument Parts Breakdown ...