Parts Replacement
Remove/Replace the Front Panel
Released: MAY 2000
Agilent 35900E Service Manual
14
Remove/Replace the Front Panel
1.
Remove the top cover. (See Figure 1.)
2.
Disconnect the front panel connector from the mother board by pressing
down on the two latches on the end of the connector on the mother board
and lifting the connector straight up. (See Figure 2)
3.
Remove the four screws holding the front panel to the unit.
4.
Gently pull the front panel forward until it is detached from the frame.
5.
Remove the ground wire connection (green wire) from the ground clip by
pulling it firmly.
6.
When you are ready to replace the front panel, reattach the green ground
wire to the ground clip on the chassis.
7.
Fit the bottom lip of the front panel under the lip of the frame and push it
against the frame, fitting the power switch into the hole.
8.
Connect the front panel cable to the connector, pressing firmly until the
ears of the connector lock over the ends of the black connector.
9.
Replace the four screws, two on each side.
Figure 2.
Removing the front panel.
Disconnect front panel
cable from connector
Disconnect ground wire
from ground clip
Ground Clip
Remove 4 screws
Summary of Contents for Agilent 35900E
Page 1: ...Agilent 35900E Dual Channel Interface Service Manual ...
Page 2: ...35900E Dual Channel Interface Agilent 35900E Service Manual ...
Page 5: ...Released MAY 2000 Agilent 35900E Service Manual 5 1 Introduction ...
Page 7: ...Released MAY 2000 Agilent 35900E Service Manual 7 2 Troubleshooting the 35900E ...
Page 10: ...Released MAY 2000 Agilent 35900E Service Manual 10 3 Parts Replacement ...
Page 18: ...Released MAY 2000 Agilent 35900E Service Manual 18 4 Instrument Parts Breakdown ...