3.5.
Environmental
3.5.1.
Environmental Specifications
Operating Location
Indoor use
Operating Temperature
0° to 50° C
Storage Temperature
−
40° to 71° C
Operating Relative Humidity
5 to 95% non-condensing
Operating Altitude
3000 m
3.5.2.
Safety
Conformity
IEC 61010-1:2001 / EN61010-1:2001 in accordance
with the European Council
Low Voltage Directive 2006/95/EC
Canada: CSA C22.2 No. 61010-1:2004
USA: UL 61010-1:2004
Installation Category II, Pollution Degree 2
Safety Class 1
3.5.3.
EMC Immunity & Emissions
Conformity
Complies with European EMC Directive 2004/108/EC
• IEC/EN 61326-2-1
• CISPR Pub 11 Group 1, class A
• AS/NZS CISPR 11
• ICES/NMB-001
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme à la norme NMB-001 du Canada.
Reference Standards
Limit
CISPR 11:2003 / EN 55011:2007
Class A, Group1
IEC 61000-4-2:2001 / EN 61000-4-2
4 kV/4 kV contact/air
IEC 61000-4-3:2002 / EN 61000-4-3
3 V/m, 80-2000 MHz, 1 V/m, 2-2.7 GHz
IEC 61000-4-4:2004 / EN 61000-4-4
0.5 kV signal lines, 1 kV power lines.
IEC 61000-4-5:2001 / EN 61000-4-5
0.5 kV line-line, 1 kV line-ground
IEC 61000-4-6:2003 / EN 61000-4-6
3 V, 0.15-80 MHz
IEC 61000-4-11:2004 / EN 61000-4-11
100% dip, 1 cycle; 60% dip, 10 cycles; 30% dip
25 cycles; 100% interruption, 250 cycles
The product was tested in a typical configuration with Agilent Technologies test systems.
User Manual: Agilent Acqiris 3-, 5-, and 8-slot cPCI Crates