484
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_read.fm
READ
“Diagram Conventions” on page 248
READ:AFANalyzer[:ALL]?
Function
Initiates and fetches an audio frequency analyzer measurement as a sequential operation.
Measurement results are returned in the following order: integrity, average audio level, average
SINAD, average audio distortion.
Query
Integrity
• Resolution: 1
Average audio level
• Range: 0 to 20 volts, 9.91 E+37
• Resolution: the lesser of 4 significant digits or 10 uV.
Average SINAD
• Range: -99.00 to 99.00 dB, 9.91 E+37
• Resolution: 0.01 dB
Average audio distortion
• Range: 0.00 to 99.00%, 9.91 E+37
• Resolution: 0.01%
READ
[:ALL]
? (returns Integrity, Rho, Frequency Error,
Time Error, Carrier Feedthrough, Phase
Error, Magnitude Error, EVM)
[:ALL]
? (returns Integrity, Frame Error Rate, Received Good
Frames, Received Blank Frames, Transmitted Good
Frames, Transmitted Blank Frames, Frame Error
Rate Confidence)