85054D
3-13
Use, Maintenance, and Care of the Devices
Connections
Using a torque wrench guarantees that the connection is not too tight, preventing possible
connector damage. It also guarantees that all connections are equally tight each time.
Prevent the rotation of anything other than the connector nut that you are tightening. It
may be possible to do this by hand if one of the connectors is fixed (as on a test port). In all
situations, however, it is recommended that you use an open-end wrench to keep the body
of the device from turning. Refer to
for part number and ordering information.
1. Position both wrenches within 90 degrees of each other before applying force. See
. Wrenches opposing each other (greater than 90 degrees apart) will cause a
lifting action which can misalign and stress the connections of the devices involved.
This is especially true when several devices are connected together.
Figure 3-5
Wrench Positions
2. Hold the torque wrench lightly, at the end of the handle only (beyond the groove). See
.
Figure 3-6
Using the Torque Wrench
Summary of Contents for 85054D
Page 5: ...85054D 1 1 1 General Information...
Page 14: ...1 10 85054D General Information When to Calibrate...
Page 15: ...2 1 2 Specifications...
Page 20: ...2 6 85054D Specifications Electrical Specifications...
Page 21: ...3 1 3 Use Maintenance and Care of the Devices...
Page 35: ...4 1 4 Performance Verification...
Page 38: ...4 4 85054D Performance Verification Recertification...
Page 39: ...5 1 5 Troubleshooting...
Page 44: ...5 6 85054D Troubleshooting Contacting Agilent...
Page 45: ...6 1 6 Replaceable Parts...
Page 48: ...6 4 85054D Replaceable Parts Introduction...
Page 49: ...A 1 A Standard Definitions...