85054D
1-9
General Information
When to Calibrate
When to Calibrate
A network analyzer calibration remains valid as long as the changes in the systematic
error are insignificant. This means that changes to the uncorrected leakages (directivity
and isolation), mismatches (source match and load match), and frequency response of the
system are small (<10%) relative to accuracy specifications.
Change in the environment (especially temperature) between calibration and
measurement is the major cause in calibration accuracy degradation. The major effect is a
change in the physical length of external and internal cables. Other important causes are
dirty and damaged test port connectors and calibration standards. If the connectors
become dirty or damaged, measurement repeatability and accuracy is affected.
Fortunately, it is relatively easy to evaluate the general validity of the calibration. To test
repeatability, remeasure one of the calibration standards. If you can not obtain repeatable
measurements from your calibration standards, maintenance needs to be performed on the
test port connectors, cables and calibration standards. Also, maintain at least one sample
of the device under test or some known device as your reference device. A verification kit
may be used for this purpose. After calibration, measure the reference device and note its
responses. Periodically remeasure the device and note any changes in its corrected
response which can be attributed to the test system. With experience you will be able to see
changes in the reference responses that indicate a need to perform the measurement
calibration again.
Summary of Contents for 85054D
Page 5: ...85054D 1 1 1 General Information...
Page 14: ...1 10 85054D General Information When to Calibrate...
Page 15: ...2 1 2 Specifications...
Page 20: ...2 6 85054D Specifications Electrical Specifications...
Page 21: ...3 1 3 Use Maintenance and Care of the Devices...
Page 35: ...4 1 4 Performance Verification...
Page 38: ...4 4 85054D Performance Verification Recertification...
Page 39: ...5 1 5 Troubleshooting...
Page 44: ...5 6 85054D Troubleshooting Contacting Agilent...
Page 45: ...6 1 6 Replaceable Parts...
Page 48: ...6 4 85054D Replaceable Parts Introduction...
Page 49: ...A 1 A Standard Definitions...