2-4
85054D
Specifications
Mechanical Characteristics
NOTE
The gages for measuring type-N connectors compensate for the designed
offset of 5.26 mm (0.207 inch), therefore, protrusion and recession readings
are in relation to a
zero
reference plane (as if the inner and outer conductor
planes were intended to be flush). Gage readings can be directly compared
with the
observed
values listed in
The pin depth value of each calibration device in this kit is not specified, but is an
important mechanical parameter. The electrical performance of the device depends, to
some extent, on its pin depth. The electrical specifications for each device in this kit take
into account the effect of pin depth on the device’s performance.
lists the typical
pin depths and measurement uncertainties, and provides observed pin depth limits for the
devices in the kit. If the pin depth of a device does not measure within the
observed
pin
depth limits, it may be an indication that the device fails to meet electrical specifications.
Refer to
for an illustration of pin depth in type-N connectors.
NOTE
When measuring pin depth, the measured value (resultant average of three
or more measurements) is
not
the true value. Always compare the measured
value with the observed pin depth limits in
to evaluate the
condition of device connectors.
Table 2-2 Pin Depth Limits
Device
Typical Pin Depth
micrometers
(10
–4
inches)
Measurement
Uncertainty
a
micrometers
(10
–4
inches)
a. Approxi2 sigma to
−
2 sigma of gage uncertainty based on studies done at the fac-
tory according to recommended procedures.
Observed Pin Depth
Limits
b
micrometers
(10
–4
inches)
b. Observed pin depth limits are the range of observation limits seen on the gage reading due
to measurement uncertainty. The depth could still be within specifications.
Opens
0 to
−
12.7
(0 to
−
5.0)
+3.8 to
−
3.8
(+ 1.5 to
−
1.5)
+3.8 to
−
16.5
(+ 1.5 to
−
6.5)
Shorts
0 to
−
12.7
(0 to
−
5.0)
+3.8 to
−
3.8
(+ 1.5 to
−
1.5)
+3.8 to
−
16.5
(+ 1.5 to
−
6.5)
Broadband
loads
0 to
−
50.8
(0 to
−
20.0)
+3.8 to
−
3.8
(+ 1.5 to
−
1.5)
+3.8 to
−
54.6
(+ 1.5 to
−
21.5)
Adapters
(7 mm end)
0 to
−
50.8
(0 to
−
20.0)
+3.8 to
−
3.8
(+ 1.5 to
−
1.5)
+3.8 to
−
54.6
(+ 1.5 to
−
21.5)
Adapters
(type-N end)
0 to
−
12.7
(0 to
−
5.0)
+3.8 to
−
3.8
(+ 1.5 to
−
1.5)
+3.8 to
−
16.5
(+ 1.5 to
−
6.5)
Summary of Contents for 85054D
Page 5: ...85054D 1 1 1 General Information...
Page 14: ...1 10 85054D General Information When to Calibrate...
Page 15: ...2 1 2 Specifications...
Page 20: ...2 6 85054D Specifications Electrical Specifications...
Page 21: ...3 1 3 Use Maintenance and Care of the Devices...
Page 35: ...4 1 4 Performance Verification...
Page 38: ...4 4 85054D Performance Verification Recertification...
Page 39: ...5 1 5 Troubleshooting...
Page 44: ...5 6 85054D Troubleshooting Contacting Agilent...
Page 45: ...6 1 6 Replaceable Parts...
Page 48: ...6 4 85054D Replaceable Parts Introduction...
Page 49: ...A 1 A Standard Definitions...