5-37
Servicing
Adjustment Procedures
Table 5-6. Required Test Equipment
Equipment
Agilent
Recommended
Model
Critical Specifications
Digital voltage meter
34401A
≥
150 Vdc range. 1 mV accuracy.
Digitizing oscilloscope
83480A with
83483A and
83485A
>10 GHz bandwidth
≤
10 ps/div amplitude scale
≥
500 MHz triggering, optical input with built-in SDH filter.
Voltage and time histograms. Built-in SONET/SDH eye mask
conformance test.
Pattern generator
70841B
0.1 to 3.3 Gb/s, 2
23
–1 PRBS.
250 mV p-p to 1200 mV p-p output.
Rise/fall times
≤
100 ps.
Clock/32 trigger.
50
Ω
output.
Clock source
70311A
16.1 MHz to 3.3 GHz, 50
Ω
, output.
<120 dBc noise
Sinewave or Squarewave, 500 mV p-p to 1500 mV p-p
Mainframe
70001A
Specific mainframe for pattern generator and clock source.
Optical power meter
—
1310/1550 nm calibration. FC connector
0.1 dB resolution, 0.2 dB accuracy
–60 to +20 dBm range
Summary of Contents for 83430A
Page 2: ...Agilent 83430A Lightwave Transmitter User s Guide...
Page 9: ......
Page 11: ......
Page 16: ...1 5 Getting Started Step 1 Inspect the Shipment...
Page 25: ......
Page 31: ...2 6 Making Measurements Using the Agilent 83430A...
Page 44: ...3 Specifications 3 3 Regulatory Information 3 6 Specifications and Regulatory Information...
Page 50: ...3 7 Specifications and Regulatory Information Regulatory Information...
Page 51: ......
Page 63: ...5 6 Servicing General Information...
Page 69: ...5 12 Servicing Troubleshooting...
Page 97: ...5 40 Servicing Adjustment Procedures Figure 5 11 Location of A3 power supply adjustments...
Page 103: ......